In situ wavelength calibration of the edge CXS spectrometers on JET

Delabie, E. and Hawkes, N. and Biewer, T. M. and O'Mullane, M. G. (2016) In situ wavelength calibration of the edge CXS spectrometers on JET. Review of Scientific Instruments, 87 (11). pp. 1-3. 11E525. ISSN 0034-6748 (https://doi.org/10.1063/1.4960306)

[thumbnail of Delabie-etal-RSI-in-situ-wavelength-calibration-of-the-edge-CXS-spectrometers-on-JET]
Preview
Text. Filename: Delabie_etal_RSI_in_situ_wavelength_calibration_of_the_edge_CXS_spectrometers_on_JET.pdf
Accepted Author Manuscript

Download (676kB)| Preview

Abstract

A method for obtaining an accurate wavelength calibration over the entire focal plane of the JET edge CXS spectrometers is presented that uses a combination of the fringe pattern created with a Fabry-Pérot etalon and a neon lamp for cross calibration. The accuracy achieved is 0.03 Å, which is the same range of uncertainty as when neglecting population effects on the rest wavelength of the CX line. For the edge CXS diagnostic, this corresponds to a flow velocity of 4.5 km/s in the toroidal direction or 1.9 km/s in the poloidal direction.

ORCID iDs

Delabie, E., Hawkes, N., Biewer, T. M. and O'Mullane, M. G. ORCID logoORCID: https://orcid.org/0000-0002-2160-4546;