A novel, contactless, portable "spot-check" device accurately measures respiratory rate
Daw, William and Kingshott, Ruth N. and Saatchi, Reza and Burke, Derek and Evans, Robert and Holloway, Alan and Travis, Jon and Jones, Anthony and Hughes, Ben and Elphick, Heather E. (2020) A novel, contactless, portable "spot-check" device accurately measures respiratory rate. Journal of Medical Devices, 14 (3). 031002. ISSN 1932-6181 (https://doi.org/10.1115/1.4046923)
Preview |
Text.
Filename: Daw_etal_JMD_2020_A_novel_contactless_portable_spot_check_device_accurately_measures_respiratory_rate.pdf
Accepted Author Manuscript License: Strathprints license 1.0 Download (720kB)| Preview |
Abstract
Abstract Respiratory rate (RR) is an important vital sign used in the initial and ongoing assessment of acutely ill patients. It is also used as a predictor of serious deterioration in a patient's clinical condition. Convenient electronic devices exist for measurement of pulse, blood pressure, oxygen saturation, and temperature. Although devices which measure RR exist, none has entered everyday clinical practice. We have developed a contactless portable respiratory rate monitor (CPRM) and evaluated the agreement in respiratory rate measurements between existing methods and our new device. The CPRM uses thermal anemometry to measure breath signals during inspiration and expiration. RR data were collected from 52 healthy adult volunteers using respiratory inductance plethysmography (RIP) bands (established contact method), visual counting of chest movements (established noncontact method), and the CPRM (new method), simultaneously. Two differently shaped funnel attachments to the CPRM were evaluated for each volunteer. Data showed a good agreement between measurements from the CPRM and the gold standard RIP, with intraclass correlation coefficient (ICC): 0.836, mean difference 0.46 and 95% limits of agreement of −5.90 to 6.83. When separate air inlet funnels of the CPRM were analyzed, stronger agreement was seen with an elliptical air inlet; ICC 0.908, mean difference 0.37 with 95% limits of agreement −4.35 to 5.08. A contactless device for accurately and quickly measuring respiratory rate will be an important triage tool in the clinical assessment of patients. More testing is needed to explore the reasons for outlying measurements and to evaluate in the clinical setting.
-
-
Item type: Article ID code: 80267 Dates: DateEvent30 September 2020Published14 May 2020Published Online8 March 2020AcceptedSubjects: Technology > Engineering (General). Civil engineering (General) > Bioengineering Department: Faculty of Engineering > Mechanical and Aerospace Engineering Depositing user: Pure Administrator Date deposited: 21 Apr 2022 15:53 Last modified: 11 Nov 2024 13:27 Related URLs: URI: https://strathprints.strath.ac.uk/id/eprint/80267