Developing an magnetic resonance-compatible catheter for cardiac catheterization
Yao, Wei and Schaeffter, Tobias and Seneviratne, Lakmal and Althoefer, Kaspar (2012) Developing an magnetic resonance-compatible catheter for cardiac catheterization. Journal of Medical Devices, 6 (4). 041002. ISSN 1932-6181 (https://doi.org/10.1115/1.4007281)
Full text not available in this repository.Request a copyAbstract
Magnetic Resonance Imaging (MRI) is a means to guide cardiac interventions and provide excellent soft tissue contrast while avoiding radiation hazards. This paper investigates and evaluates a new Magnetic Resonance (MR)-compatible catheter for cardiac catheterization. Important mechanical properties of the catheter are measured and investigated; these include flexibility, pushability, and torquability. The mechanical performance of the MR-compatible and steerable catheter is benchmarked against three commercially-available clinical ablation catheters that are not MR-compatible. The MR-compatibility of the proposed catheter is also evaluated through an experimental study inside a 1.5 T MRI scanner. The new catheter is shown to have a mechanical performance comparable to that of existing catheters while being MR compatible.
-
-
Item type: Article ID code: 40850 Dates: DateEventDecember 2012PublishedSubjects: Medicine > Other systems of medicine Department: Faculty of Engineering > Bioengineering Depositing user: Pure Administrator Date deposited: 09 Aug 2012 15:20 Last modified: 11 Nov 2024 10:12 Related URLs: URI: https://strathprints.strath.ac.uk/id/eprint/40850