On-line detection and quantification of trace impurities in vaporisable samples by direct liquid introduction process mass spectrometry

Owen, Andrew W. and Nordon, Alison and Littlejohn, David and Lynch, Thomas P. and Lancaster, J. Steven and Wright, Robert G. (2014) On-line detection and quantification of trace impurities in vaporisable samples by direct liquid introduction process mass spectrometry. Analytical Methods, 6 (20). pp. 8148-8153. ISSN 1759-9660 (https://doi.org/10.1039/c4ay01064g)

[thumbnail of On-line detection and quantification of trace impurities in vaporisable samples by direct liquid introduction process mass spectrometry]
Preview
PDF. Filename: Online_detection_of_BTX_revised_version_text_and_figs.pdf
Accepted Author Manuscript

Download (440kB)| Preview

Abstract

A thermal vaporiser has been designed for analysis of liquid streams by a process mass spectrometer normally used for gas analysis. Concentrations of benzene, toluene and o-xylene at mg kg-1levels in ethanol were determined from continuous vaporisation of the liquid. Ions with m/z values of 39, 57, 73, 77, 78, 91, 92 and 106 were selected and the optimal regression model (multiple linear regression with mean-centring) was found using an automated design of experiments approach to calibration model selection. It was discovered that the linearity of the response allowed excellent calibration to be performed using only four standards (at 0 and 110 mg kg-1for each of the three analytes) and that there were minimal inter-analyte interferences. The detection limit of benzene, toluene and o-xylene was 0.5, 0.8 and 0.5 mg kg-1, respectively. Average differences between the actual and predicted concentrations, expressed as a percentage of the actual concentrations, for 27-82 mg kg-1of benzene, toluene and o-xylene were 0.5-1.4%, 0.0-0.4% and 0.3-1.6%, respectively, while the average relative standard deviations were 1.3-2.6%, 1.0-2.5% and 1.1-2.3%, respectively. Detection of 3 mg kg-1changes in the concentration of each of the analytes (at the 36 mg kg-1level) was also demonstrated, indicating the sensitivity of the technique and the potential ability of the procedure to detect minor deviations in the specification of process streams from continuous analysis.