Comparison of the results of 4 users of a contemporary CADCAM system

McGarry, T. and McHugh, B. (2007) Comparison of the results of 4 users of a contemporary CADCAM system. Prosthetics and Orthotics International, 31 (1). pp. 27-35. ISSN 1746-1553 (https://doi.org/10.1080/03093640600942101)

[thumbnail of strathprints015048.pdf]
Preview
PDF. Filename: strathprints015048.pdf
Accepted Author Manuscript

Download (79kB)| Preview

Abstract

The objective of this study was to investigate the variation of measurements recorded when four different users of the Tracer CAD system trace a model of known dimensions and volume. This complements a previous study by the same authors Mc Garry and Mc Hugh, (2005) where the accuracy and consistency of a single user was measured. Landmarks were added to indicate proximal, distal, anterior, medial and lateral regions of a specially manufactured cylindrical nylon 6.6 model. Four circumferential lines were added at regular intervals along the length of the cylinder with a view to calculating diameters and volumes relative to these landmarks. The model was measured using a comparator with guaranteed accuracy to one hundredth of a millimetre, and was traced using the Tracer CAD system by four different users. The difference in mean volume between measured results and Tracer CAD scans of differing users ranged to -3%. Individual trace volumes varied by up to -7.85%. In all volumes measured, eleven out of twelve maximum volume percentage differences measured greater than 2 %, and of these, seven results showed maximum volume percentage difference to measure greater than 4%.

ORCID iDs

McGarry, T. ORCID logoORCID: https://orcid.org/0000-0002-0738-5906 and McHugh, B.;