Impact of temperature on ophthalmic viscosurgical devices and clinical implications : a pilot study
Henry, Caoimhe and Castro, Alejandro Thompson and Belkessa, Mohamed and Rycroft, Ewan and Syntouka, Ioanna M. and Silva, Andreia and Sillars, Fiona and Oliveira, Mónica S.N. and Lockington, David (2024) Impact of temperature on ophthalmic viscosurgical devices and clinical implications : a pilot study. Journal of Cataract and Refractive Surgery, 50 (11). pp. 1173-1179. ISSN 1873-4502 (https://doi.org/10.1097/j.jcrs.0000000000001513)
Text.
Filename: Henry-etal-2024-JCRS-The-impact-of-temperature-on-Ophthalmic-Viscoelastic-Devices-and-clinical-implications.pdf
Accepted Author Manuscript Restricted to Repository staff only until 2 July 2025. Download (2MB) | Request a copy |
Abstract
Purpose: To evaluate how temperature affects the rheology of common ophthalmic viscosurgical devices (OVDs) and clinical implications. Setting: Tennent Institute of Ophthalmology, Glasgow, with Department of Mechanical and Aerospace Engineering, and Advanced Materials Research Laboratory, University of Strathclyde, Glasgow, United Kingdom. Design: Laboratory pilot study. Methods: The viscous and elastic responses of 3 OVDs (Eyefill-SC/Eyefill-C/Eyefill-HD) were measured using rotational and extensional rheometers at clinically relevant temperatures (5°C, 25°C, 37°C). Thermal properties were evaluated using differential scanning calorimetry and laser-flash analysis. Results: The OVDs tested exhibited viscoelastic properties and shear-thinning behavior. Apparent viscosities and relaxation time were higher at lower temperatures. The Eyefill-C and Eyefill-SC exhibited predominantly viscous character at low frequencies with a transition to predominantly elastic behavior at high frequencies. An increase in temperature led to a decrease in relaxation time under shear and extension. At low frequencies, Eyefill-C and Eyefill-SC moduli increase with decreasing temperatures. Eyefill-HD at 25°C and 37°C displays 2 crossover points, with the storage modulus dominating at low and high frequencies indicating a predominantly elastic behavior. Thermal property analysis revealed Eyefill-C had the lowest thermal conductivity. Conclusions: This pilot study confirms our clinical experience that OVD properties are affected by low temperatures, with increased viscosities at low shear rates and higher relaxation times. Cold OVD can cause greater resistance to initiation of intraocular lens (IOL) injection system forces (compared with warmer OVD). Excessively forced injection using cold OVD could contribute to inadvertent cannula detachment if incorrectly assembled, or uncontrolled IOL release leading to avoidable injury.
ORCID iDs
Henry, Caoimhe, Castro, Alejandro Thompson, Belkessa, Mohamed, Rycroft, Ewan, Syntouka, Ioanna M., Silva, Andreia, Sillars, Fiona, Oliveira, Mónica S.N. ORCID: https://orcid.org/0000-0002-1836-4692 and Lockington, David;-
-
Item type: Article ID code: 90563 Dates: DateEvent30 November 2024Published2 July 2024Published Online24 June 2024Accepted23 October 2023SubmittedSubjects: Technology > Mechanical engineering and machinery Department: Faculty of Engineering > Mechanical and Aerospace Engineering
Strategic Research Themes > Ocean, Air and Space
Technology and Innovation Centre > Advanced Engineering and ManufacturingDepositing user: Pure Administrator Date deposited: 13 Sep 2024 14:37 Last modified: 30 Nov 2024 04:01 URI: https://strathprints.strath.ac.uk/id/eprint/90563