Ultrasound mediated delivery of quantum dots from a proof of concept capsule endoscope to the gastrointestinal wall
Stewart, Fraser and Cummins, Gerard and Turcanu, Mihnea V. and Cox, Benjamin F. and Prescott, Alan and Clutton, Eddie and Newton, Ian P. and Desmulliez, Marc P.Y and Thanou, Maya and Mulvana, Helen and Cochran, Sandy and Näthke, Inke (2021) Ultrasound mediated delivery of quantum dots from a proof of concept capsule endoscope to the gastrointestinal wall. Scientific Reports, 11. 2584. ISSN 2045-2322 (https://doi.org/10.1038/s41598-021-82240-1)
Preview |
Text.
Filename: Stewart_etal_SR_2021_Ultrasound_mediated_delivery_quantum_dots_proof_of_concept_capsule_endoscope.pdf
Final Published Version License: Download (1MB)| Preview |
Abstract
Biologic drugs, defined as therapeutic agents produced from or containing components of a living organism, are of growing importance to the pharmaceutical industry. Though oral delivery of medicine is convenient, biologics require invasive injections because of their poor bioavailability via oral routes. Delivery of biologics to the small intestine using electronic delivery with devices that are similar to capsule endoscopes is a promising means of overcoming this limitation and does not require reformulation of the therapeutic agent. The efficacy of such capsule devices for drug delivery could be further improved by increasing the permeability of the intestinal tract lining with an integrated ultrasound transducer to increase uptake. This paper describes a novel proof of concept capsule device capable of electronic application of focused ultrasound and delivery of therapeutic agents. Fluorescent markers, which were chosen as a model drug, were used to demonstrate in vivo delivery in the porcine small intestine with this capsule. We show that the fluorescent markers can penetrate the mucus layer of the small intestine at low acoustic powers when combining microbubbles with focused ultrasound during in vivo experiments using porcine models. This study illustrates how such a device could be potentially used for gastrointestinal drug delivery and the challenges to be overcome before focused ultrasound and microbubbles could be used with this device for the oral delivery of biologic therapeutics.
ORCID iDs
Stewart, Fraser, Cummins, Gerard, Turcanu, Mihnea V., Cox, Benjamin F., Prescott, Alan, Clutton, Eddie, Newton, Ian P., Desmulliez, Marc P.Y, Thanou, Maya, Mulvana, Helen ORCID: https://orcid.org/0000-0002-5058-0299, Cochran, Sandy and Näthke, Inke;-
-
Item type: Article ID code: 75248 Dates: DateEvent28 January 2021Published14 January 2021AcceptedSubjects: Medicine > Pharmacy and materia medica Department: Faculty of Engineering > Electronic and Electrical Engineering
Faculty of Engineering > Biomedical EngineeringDepositing user: Pure Administrator Date deposited: 02 Feb 2021 15:59 Last modified: 11 Nov 2024 12:57 URI: https://strathprints.strath.ac.uk/id/eprint/75248