Performance of photovoltaic arrays in-vivo and characteristics of prosthetic vision in animals with retinal degeneration

Lorach, Henri and Goetz, Georges and Mandel, Yossi and Lei, Xin and Kamins, Theodore I. and Mathieson, Keith and Huie, Philip and Dalal, Roopa and Harris, James S and Palanker, Daniel (2015) Performance of photovoltaic arrays in-vivo and characteristics of prosthetic vision in animals with retinal degeneration. Vision Research, 111 (B). pp. 142-148. ISSN 0042-6989 (https://doi.org/10.1016/j.visres.2014.09.007)

[thumbnail of Lorach-etal-VR-2015-Performance-of-photovoltaic-arrays-in-vivo]
Preview
PDF. Filename: Lorach_etal_VR_2015_Performance_of_photovoltaic_arrays_in_vivo.pdf
Accepted Author Manuscript
License: Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 logo

Download (1MB)| Preview

Abstract

Loss of photoreceptors during retinal degeneration leads to blindness, but information can be reintroduced into the visual system using electrical stimulation of the remaining retinal neurons. Subretinal photovoltaic arrays convert pulsed illumination into pulsed electric current to stimulate the inner retinal neurons. Since required irradiance exceeds the natural luminance levels, an invisible near-infrared (915nm) light is used to avoid photophobic effects. We characterized the thresholds and dynamic range of cortical responses to prosthetic stimulation with arrays of various pixel sizes and with different number of photodiodes. Stimulation thresholds for devices with 140μm pixels were approximately half those of 70μm pixels, and with both pixel sizes, thresholds were lower with 2 diodes than with 3 diodes per pixel. In all cases these thresholds were more than two orders of magnitude below the ocular safety limit. At high stimulation frequencies (>20Hz), the cortical response exhibited flicker fusion. Over one order of magnitude of dynamic range could be achieved by varying either pulse duration or irradiance. However, contrast sensitivity was very limited. Cortical responses could be detected even with only a few illuminated pixels. Finally, we demonstrate that recording of the corneal electric potential in response to patterned illumination of the subretinal arrays allows monitoring the current produced by each pixel, and thereby assessing the changes in the implant performance over time.