Spatial resolution of the Medipix2 X-ray imaging detector system

Davidson, D.W. and Tlustos, L and Mathieson, K and O'Shea, V and Smith, K.M. and Rahman, M; Metzler, S.D., ed. (2004) Spatial resolution of the Medipix2 X-ray imaging detector system. In: 2003 IEEE Nuclear Science Symposium Conference Record. IEEE NUCLEAR SCIENCE SYMPOSIUM - CONFERENCE RECORD, 3 . IEEE, pp. 2727-2729. ISBN 0-7803-8257-9 (https://doi.org/10.1109/NSSMIC.2003.1352450)

Full text not available in this repository.Request a copy

Abstract

The charge sharing present in a 55 micrometer square pixel matrix of 300 micrometer thick silicon is examined. This is achieved using a commercial simulation package to investigate the diffusion effects on the charge cloud of created electrons and holes. The total fraction of charge shared is found to be 20.8%, but this can be reduced by the addition of a minimum threshold level of detection. The optimum level giving all of the signal but with the minimum of noise is found to leave 10.4% charge shared on average.